Alignment and Measurement
Intensity too low:
A weak autocorrelation signal uses only a minor part of the given dynamic range. This can lead
to an erroneous measurement of the FWHM value due to the low signal-to-noise ratio (see Figure
4.19).
Figure 4.19.: A too weak signal leading to a low dynamic is indicated by the info box
Action: Increase "Sensitivity" or laser power!