B.
TCP/IP Command Set
scoutlink:SYSTEM:MOTOR?
scoutlink:SYS:MOTOR?
Get Motor Type
<version>
Motor Type as string,
Example:
scoutlink:sys:motor?
scoutlink:SYSTEM:HELP?
scoutlink:SYS:HELP?
Get List of all SCPI Commands
<command_list>
Command List as array of s in block data format,
Example:
scoutlink:sys:help?
scoutlink:STATUS:AVERAGE<number>
scoutlink:STA:AVERAGE<number>
Set number of measurements used for averaging
<number>
Number of Measurements 0: Averaging OFF, 1: 2 Measurements, 2: 4 Mea-
surements, 3: 8 Measurements, 4: 16 Measurements as integer, range: 0 ... 4,
Example:
scoutlink:sta:average 1
scoutlink:STATUS:AVERAGE?
scoutlink:STA:AVERAGE?
Get number of measurements used for averaging
<number>
Number of Measurements 0: Averaging OFF, 1: 2 Measurements, 2: 4 Mea-
surements, 3: 8 Measurements, 4: 16 Measurements as integer, range: 0 ... 4,
Example:
scoutlink:sta:average?
scoutlink:STATUS:CLEAR_DISPLAY
scoutlink:STA:CLRD
Reset the Averaging function and clear the current Fit
scoutlink:STATUS:FITTYPE<type>
scoutlink:STA:FITTYPE<type>
Set type of curve t to apply to measured ACF
<type>
Fittype 0/OFF/NONE: No Curve tting performed, 1/GAUSSIAN: Fit Gaussian
Model, 2/SECH2: Fit Sech2 Model, 3/LORENTZ: Fit Lorentz Model as integer, range: 0 ... 3,
Pulse
Scout2 Version 1.1
55
July 2015