B.
TCP/IP Command Set
Example:
scoutlink:mot:sfrna?
scoutlink:MOTOR:SCANRANGE<scanrange>
scoutlink:MOT:SCR<scanrange>
Set ScanRange
<scanrange>
Scanrange 0/ZERO: Zeroscan, 1/150: 150 fs, 2/500: 500 fs, 3/1500: 1.5
ps, 4/5000: 5 ps, 5/15000: 15 ps, 6/30000: 30 ps (optional) as integer, unit: fs
Example:
scoutlink:mot:scr 15000
scoutlink:MOTOR:SCANRANGE?
scoutlink:MOT:SCR?
Get ScanRange
<scanrange>
Scanrange 0: Zeroscan, 150: 150 fs, 500: 500 fs, 1500: 1.5 ps, 5000: 5
ps, 15000: 15 ps, 30000: 30 ps (optional), as integer, unit: fs
Example:
scoutlink:mot:scr?
scoutlink:DETECTOR:GAIN<value>
scoutlink:DET:GAIN<value>
Set PMT Gain value (Function has no effect, if standard optics units (w/o a PMT) is
connected)
<value>
PMT Gain Value as integer, range: 300 ... 1000,
Example:
scoutlink:det:gain 450
scoutlink:DETECTOR:GAIN?
scoutlink:DET:GAIN?
Get PMT Gain value
<value>
PMT Gain Value as integer, range: 300 ... 1000,
Example:
scoutlink:det:gain?
scoutlink:DETECTOR:SENSITIVITY<number>
scoutlink:DET:SEN<number>
Set Sensitivity
<number>
Detector Sensitivity 1: Low Sensitivity, 10: High Sensitivity, 100: (optional
"HighSen"-Feature) as integer,
Pulse
Scout2 Version 1.1
58
July 2015