4.
Alignment and Measurement
Scan range too large:
If the chosen scan range is too large with respect to the detected width of the autocorrelation
trace, the data resolution may be too low for a precise determination of the ACF value (see Figure
4.17). Note that an error message indicating a too large scan range will appear in the Info box.
This is not the case in the Figure below.
Figure 4.17.: The Scan Range is too large. The ACF is comprised by only few data points. This may
lead to an unprecise measurement.
Action: Decrease the scan range!
Pulse
Scout2 Version 1.1
37
July 2015