February 2019
14
QUANTX-300
Specifications
PERFORMANCE SPECIFICATIONS
Wavelength Range
1
325 – 1800 nm
Nominal Spot Size
2
1.1 x 1.2mm ±0.1mm
Nominal Working Distance
3
85mm ± 1mm
Spectral Response Repeatability
4
+/- 0.5% of measured value
Spectral Response Uncertainty
4
+/- 3% of measured value, 325 – 1590 nm
+/- 8% of measured value, 1600 – 1800 nm
EQE Uncertainty
4
+/- 3% of measured value, 325 – 1590 nm
+/- 8% of measured value, 1600 – 1800 nm
IQE Uncertainty
4,6
+/- 4.5%, 350 – 1800 nm
Specular Reflectance Uncertainty
4
+/- 2.1 % of measured value up to 20% reflective surfaces,
350 – 1800 nm
Specular Reflectance
Repeatability
4
+/- 1.0 % of measured value
Spectral Bandwidth (FWHM)
5
1 - 40 nm
Wavelength Accuracy
7
±0.5 nm
Polarization of beam at sample
2:1 linear, S-polarized
Detector Type
8
Summed Silicon/Germanium bounce detector
Measurement Type
Chopped beam AC measurement with Virtual Lock-In
Amplifier
Measurement Voltage Range
+/- 10V
Optical Chopper Frequency
4.0 Hz – 100.0Hz
Chopper Frequency Resolution
0.1 Hz
Sample DC Gain Settings
9
10, 100, 1k, 10k, 100k, 1M, 10M
AC Gain Settings
1, 10, 100, 1000
Sample Current Range
10nA – 1.0A
Voltage Bias
+/- 10 V up to 0.5 A
Voltage Bias Resolution
0.1 V
ELECTRICAL / MECHANICAL SPECIFICATIONS
Compliance CE
compliance
Size (H x W x D)
0.54 x 0.85 x 0.40 m (21.1 x 33.4 x 14.6 in.)
Weight
70.0 lbs (31.8 kg)
Power Requirements
100 - 115 VAC,3 Amps, 48 – 66 Hz
220 - 240 VAC, 1.5 Amps, 48 – 66 Hz
Temperature
23 ± 5 °C operating; -40 to +70
C storage
Humidity
< 85 % relative humidity, non-condensing
NOTES
1.
Can scan from 300 – 1800 with reduced accuracy and repeatability between 300-330nm.
2.
Over wavelength range of 325 nm to 1800 nm.
3.
Working distance measured from bottom of large flange containing output lens to top of sample.
4.
Uncertainty with k=2 confidence. Value is a percent of the measured value.
5.
Slit to bandwidth conversion for monochromator as configured with 600 lines/mm gratings is 13.3nm/1mm slit width.
6.
Excludes Diffuse measurement for standard product.
7.
As specified by Newport for CS130B over narrow wavelength range. May be larger uncertainty near low and high ends of
wavelength range.
8.
Bounce detector patent pending
9.
DC detector gain setting for Monitor, Reflectance, and Diffuse detectors limited to 10K, 100K, 1M, 10M.