February 2019
50
QUANTX-300
you to select the wavelength the monochromator switches from grating #1 to grating #2 for higher
efficiency. Grating #1 has high enough efficiency through the scan range from 300-1100nm so no
switch is needed during this wavelength range. As a result, the switch over wavelength is set to a
default of 1110nm for the IQE-200B, so no switch takes place. Grating #2 will be used to scan from
1110 to 1800nm.
Function
Sets the switch-over wavelength for
the diffraction gratings
Grating #1 is efficient enough to scan the
full range up to 1100nm so no switch over
is needed. .
Range
325– 1800nm
Typical = 1110nm
Increment/
decrement
1nm
Use arrows to increase/decrease
wavelength. Press enter to set wavelength
to new value.
Default
1110nm