February 2019
82
QUANTX-300
3. The Cornerstone 130B should be automatically detected and the program should initialize.
4. We will use the peak found at 824nm to correct the wavelength calibration of the
monochromator.
5. Input “824” into the “Set Wavelength” field of the monochromator utility and press <Enter> to
send the monochromator to that wavelength.