February 2019
83
QUANTX-300
6. Next, click the “Calibration” button.
7. Since the peak found at 824nm should actually be at 823nm, input this into the “Calibrate
Wavelength” field of the Calibration Settings popup window and press <Enter>. Press “OK”
in the warning message box that appears.