Instructions for the user
56
ENG_Manual_NW30-R01-01
Sterilizer NEWSTER
NW30
Then close the inverter group board and turn the main switch to the “1” (on) position. Finally,
follow the instructions at the bottom of this list of alarms.
If the display shows
THERMAL PROTECTION FUMES SUCTION
turn off the power, positioning the main switch in the “0” (off) position, open the electric
board and find the breaker marked MT2 and move its lever to the “1” (on) position. Then close
the switchboard and turn the main switch in the “1” (on) position. Finally, follow the
instructions at the bottom of this list of alarms.
If the display shows
THERMAL PROTECTION FILTER SUCTION MOTOR
turn off the power, positioning the main switch in the “0” (off) position, open the electric board
and find the breaker marked MT1 and move its lever to the “1” (on) position. Then close the
electic board and turn the main switch to the “1” (on) position. Finally, follow the instructions
at the bottom of this list of alarms.
If the display shows
CHECK MT11 PROTECTION OF MAINMOTOR SERVO FAN
turn off the power, positioning the main switch in the “0” (off) position, open the electric board
and find the breaker marked MT11 and move its lever to the “1” (on) position. Then close the
electic board and turn the main switch to the “1” (on) position. Finally, follow the instructions
at the bottom of this list of alarms.
If the display shows
LOW WATER PRESSURE
check that both the water tap is open and that there are no obstructions along the hose that
carries the water to the machine. Then follow the instructions at the bottom of this list of
alarms.
If the display shows
WATER PRESSURE RECIRCULATING SYSTEM TOO LOW
please check that the water recirculating system pump is working.
If the water recirculating system pump is working first of all you have to check if there is
enought water in the circuit and, if necessary, refill with other water.
If the problem remains contact the assistance service.
Summary of Contents for NW30
Page 1: ...ITA_MANUALE_NW30 R01 01 Sterilizzatore di rifiuti sanitari a rischio infettivo Manuale NW30 ...
Page 19: ...Sezione sicurezza 19 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW50 2 Sezione sicurezza ...
Page 45: ...Sezione funzionamento 45 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 27 ...
Page 90: ...Sezione manutenzione 90 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 54 ...
Page 92: ...Sezione manutenzione 92 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 55 ...
Page 94: ...Sezione manutenzione 94 ITA_MANUALE_NW30 R01 01 Sterilizzatore NEWSTER NW30 Fig 56 ...
Page 125: ...ENG_Manual_NW30 R01 01 Healthcare risk waste sterilizer User Manual NW30 ...
Page 129: ...General information 5 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 1 General information ...
Page 140: ...Safety notes 16 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 2 Safety notes ...
Page 144: ...Safety notes 20 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 17 ...
Page 159: ...Instructions for the user 35 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 26 ...
Page 188: ...Maintenance 64 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 4 Maintenance ...
Page 191: ...Maintenance 67 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 55 ...
Page 193: ...Maintenance 69 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 56 ...
Page 195: ...Maintenance 71 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW30 Fig 57 ...
Page 215: ...Trouble shooting 91 ENG_Manual_NW30 R01 01 Sterilizer NEWSTER NW50S 5 Trouble shooting ...