18
19
3.2. Assembly of the NSZ818 zoom stereo microscope
The diagram below shows the sequence of assembly of the various modules.
The numbers indicate the order of assembly.
When assembling the microscope, make
sure that all parts are free of dust and dirt and
avoid scratching any parts or touching glass
surfaces.
Please use the three hexagonal screwdriv-
ers for assembling and replacing components
3.2.1. Detailed assembly procedure
1
Mounting the stand column
Illustration 004: Mounting the stand column.
V
II
IV
III
I
BEFORE YOU START:
Select a lo-
cation free of vibration.
Put the stand base (I) flat on the
workbench.
Align the connecting plate (II) of the
stand column (III) with the mounting
plate (IV) of the stand base (I).
Use the four supplied hexagon socket
head screws (V) to connect the stand
base and the stand column firmly.
USE
the supplied hexagonal
screwdriver for assembling.
Check input voltage: The input voltage and
supply voltage indicated on the subplate of the
microscope must be consistent, otherwise the
microscope will be seriously damaged.
3
DESIGN OF TRINO ZOOM STEREO MICROSCOPE
3
DESIGN OF TRINO ZOOM STEREO MICROSCOPE
7
3
4
5
2
1
6
Illustration 003: Assembly of zoom stereo microscope NSZ818.
Three hexagonal screwdivers.