4. Advanced Operation
PROGRAMMABLE AC POWER SOURCE
104
Initial Step
Initial Step is assigned to the stand-by state before the simulation starts. The DP Series also moves
to the Initial Step and becomes stand-by state after the simulation ends, except when the repeat
count is specified to 0 (= infinite count).
Transition Step (Trans 1 and Trans 2)
The Step Times of Trans 1 and Trans 2 can be set to zero for simulating the abrupt voltage and
phase changes. When the Step Time is set to the value other than zero, the set voltage and frequency
are swept linearly at the previous and next steps. Note that the AC voltage, frequency, start phase,
and Stop Phase cannot be set in the Transition Steps.
Other issues
The Exit operation is available during the simulation. Performing the Exit operation moves to
the Initial Step. The output remains on.
You can set the repeat on/off and repeat count for the simulation. When the repeat is set to on
and the repeat count to zero (0), the testing repeats infinitely.
--------
Notes
----------------------------------------------------------------------------------------------------------------
The Simulation function does not support the standard tests defined by organizations such as
IEC. Use this function for a preliminary testing.
---------------------------------------------------------------------------------------------------------------------------------
4.3.2 Simulation Function Parameters
Table 4-4 shows the parameters of Simulation function. The common parameters are common to one
simulation. In the Simulation function, the signal source is fixed to INT sine wave and the ACDC
mode to ACDC mode. When in polyphase, the balanced mode is always used. In each step, set the
step-control parameter and the intra-Step parameter. As for intra-Step parameter, the range that can
be set differ depending on the output range (100 V/200 V range) as in the case of continuous output
function.
Table 4-4 Simulation Function Parameters
Common
parameter
Step-control parameter
Intra-Step parameter
Output range
AC/DC mode
(Fixed to ACDC)
Waveform
(Fixed to sine
wave)
Step Time
Start Phase
Stop Phase
Synchronous step output (2bit)
Trigger output
Repeat count (1 to 9999 or infinite)
Frequency
AC voltage
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......