4.3 Using Power Fluctuation Testing (Simulation) Function
DP Series
125
2.
In [PwOn Mode] item, Select [3: Sim].
--------
Notes
----------------------------------------------------------------------------------------------------------------
To set so as to continuous output function is selected at power-on, select [1: Cont] in [PwOn
Mode] item.
---------------------------------------------------------------------------------------------------------------------------------
4.3.11 Simulation Control by External Control
External control enables to start/stop a simulation and to recall the memory. For details, see
4.3.12 Screen Overview
The Simulation function screen is broadly divided to the Simulation Edit View and the Simulation
Control View. The Simulation Edit View opens when you select the power fluctuation testing
(Simulation) icon from the root menu. After pushing [Compile] soft-key in Simulation Edit View,
the screen transitions to the Simulation Control View.
--------
Notes
----------------------------------------------------------------------------------------------------------------
You cannot transition to root menu by pushing the MENU key in the Simulation Control View.
To transition to the root menu, first move to the Simulation Edit View and then push the
MENU key.
---------------------------------------------------------------------------------------------------------------------------------