4. Advanced Operation
PROGRAMMABLE AC POWER SOURCE
164
--------
Notes
----------------------------------------------------------------------------------------------------------------
When the Autocal function is on, the calibration factor is always used to perform the
calibration as a multiplier.
If the conditions for setting the [Acal] to [ON] shown in
are not satisfied, the Autocal
function cannot be turned on. However, if any one of the conditions required for setting [Acal]
to [ON] becomes disabled while the Autocal function is on, the ON setting of the Autocal
function is retained.
If the difference is out of the "Calibration range" shown in
, the Autocal function is
forcibly set to off
If the limiter operates while the Autocal function is on, the Autocal calibration is still applied,
but the limit is applied on the output. If the protection function is activated, the output is
turned off, but the Autocal setting remains on.
When the output range is changed, the Autocal function is forcibly set to off.
When the Autocal function is forcibly set to off, the settings saved in the System Setting
Memory are not updated.
The Autocal function continues to use the calibration factor obtained when Autocal is turned
on until Autocal is turned off. In contrast, the AGC function updates the calibration factor
continuously while the AGC function is on. Therefore, in the Autocal function, the output
voltage may not be calibrated properly when the load is fluctuated, but once Autocal is turned
on, there is no response time needed for calibration. Meanwhile, in the AGC function, the
output voltage is calibrated properly even when the load is fluctuated, but there is the response
time to reflect the update of the calibration factor.
---------------------------------------------------------------------------------------------------------------------------------
Turning on the Autocal function
1.
Turn on the output.
2.
Open the AGC/Autocal set window. Following two methods are available.
(a)
Shortcut operation:
+
(b)
Push the [Misc] soft-key and select [1: AGC/Auto Cal].
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......