8. Troubleshooting
PROGRAMMABLE AC POWER SOURCE
232
Table 8-3 Actions When a Failure is Suspected (Continued)
Category
Symptom
Possible Cause
Necessary Actions
Limiter/setti
ng range
limitation-
related
problem
The output
current
peak-hold value
(measured) is
larger than the
peak current
limiter setting
value.
The output current
overshoot occurred for
the limit value.
While the peak current limiter is
activated, the overshoot may
occur depending on the load.
→
The output mode is
incorrect.
The waveform setting is
incorrect.
Remote
sensing
problem
The Remote
sensing function
cannot be turned
on.
The mode is not the
continuous output mode.
The Remote sensing function can
be set to on in the AC-INT,
AC-VCA, AC-SYNC, DC-INT,
and DC-VCA modes. However,
for the AC mode, it can be set to
on only when the waveform is a
sine wave. It is turned off when
the Sequence or Simulation
function is selected.
→
The output mode is
incorrect.
The waveform setting is
incorrect.
The mode is not the
continuous output mode.
The AGC function can be set to
on in the AC-INT, AC-VCA,
AC-SYNC, DC-INT, and
DC-VCA modes. However, for
the AC mode, it can be set to on
only when the waveform is a sine
wave. It is turned off when the
Sequence or Simulation function
is selected.
→
Problem of
AGC
function
The AGC
function cannot
be turned on.
The Autocal function is
on.
When the Autocal function is on,
the AGC function cannot be used.
Turn off the Autocal function.
→
The output mode is
incorrect.
The waveform setting is
incorrect.
The mode is not the
continuous output mode.
The Autocal function can be set to
on in the AC-INT, AC-VCA,
AC-SYNC, DC-INT, and
DC-VCA modes. However, for
the AC mode, it can be set to on
only when the waveform is a sine
wave. It is turned off when the
Sequence or Simulation function
is selected.
→
Problem of
Autocal
function
The Autocal
function cannot
be turned on.
The AGC function is on. When the AGC function is on, the
Autocal function cannot be used.
Turn the AGC function off.
→
Problem of
remote
control
See "Instruction Manual (Remote Control)" in the attached CD-ROM.
The external control
function is disabled.
Enable the external control
function.
→
Problem of
external
control
function
The product
cannot be
controlled by the
external control
function.
The Remote state is set
by the communication
interface.
The external control function is
disabled in the Remote state.
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......