10. Specifications
PROGRAMMABLE AC POWER SOURCE
256
RMS current limiter
Single-phase model, Polyphase model, Multi-phase model
015S,030D,045T
045M (polyphase
output)
030S,060D,090T
090M (polyphase
output)
045S, 090D
045M
(single-phase
output)
060S, 120D
Setting range
(effective value)
0.8 A
to
15.8 A /
0.8 A
to
7.9 A
1.5 A
to
31.5 A /
1.5 A
to
15.8 A
2.3 A
to
47.3 A /
2.3 A
to
23.7 A
3.0 A
to
63.0 A /
3.0 A
to
31.5 A
Factory default
15.8 A / 7.9 A
31.5 A / 15.8 A
47.3 A / 23.7 A
63.0 A / 31.5 A
Resolution 0.1
A
Limiter operation
Select whether to recover automatically (continuous, factory default) or
turn the output off when the limited state continues over the specified time
(1 s to 10 s, resolution 1 s).
Single-phase model and Single-phase output of Multi-phase model
075S 090S,
090M 105S 120S
Setting range
(effective value)
3.8 A
to
78.8 A /
3.8 A
to
39.4 A
4.5 A
to
94.5 A /
4.5 A
to
47.3 A
5.3 A
to
110.3 A /
5.3 A
to
55.2 A
6.0 A
to
126.0 A /
6.0 A
to
63.0 A
Factory default
78.8 A / 39.4 A
94.5 A / 47.3 A
110.3 A / 55.2 A
126.0 A / 63.0 A
Resolution 0.1
A
Limiter operation
Select whether to recover automatically (continuous, factory default) or
turn the output off when the limited state continues over the specified time
(1 s to 10 s, resolution 1 s).
Note: If you increased or decreased the number of units by the power unit energization setting, the
factory
default
setting corresponding to the capacity is used. (Example: In the case that only
3 units operate in 6 kVA model, the factory default setting of 4.5 kVA model is used.)
10.14 Setting Range Limit Function
This is the limit function for the setting of the internal signal source. It works when the signal source
is INT, VCA (frequency setting limit only), SYNC (voltage setting limit only), or ADD (internal
signal source only). The limitation does not work for the Sequence and Simulation. It does not also
work for the external signal source of EXT and ADD.
In the polyphase model, polyphase system, and polyphase output of the Multi-phase model, the
setting is common to all the phases.
Voltage setting limit 1 (in the AC mode, and sine wave or clipped sine wave is selected)
All
models
Phase voltage setting
7.0 V to 155.0 V / 14.1 V to 310.0 V
Line voltage setting (single-phase three-wire)
14.0 V to 310.0 V / 28.2 V to 620.0 V
Setting range
(effective value)
Line voltage setting (three-phase four-wire)
12.2 V to 268.4 V / 24.4 V to 536.8 V
Factory default
Phase voltage setting, 155.0 V/310.0 V
Resolution
Phase voltage setting: 0.1 V, line voltage setting: 0.2 V
Note: The line voltage setting is available only when the output voltage setting is set as the line
voltage and sine wave is selected in the balanced mode of the polyphase model, polyphase
system, and polyphase output of the Multi-phase model.
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......