10.15 Remote Sensing
DP Series
257
Voltage setting limit 2 (other than Voltage setting limit 1, phase voltage setting only)
All
models
Setting Range
(Peak value)
+10.0 V to +220.0 V / +20.0 V to +440.0 V
Positive
voltage
Factory
default
+220.0 V / +440.0 V
Setting Range
(Peak value)
−
220.0 V to
−
10.0 V /
−
440.0 V to
−
20.0 V
Negative
voltage
Factory
default
−
220.0 V /
−
440.0 V
Resolution 0.1
V
Note: The limitation is applied to the additional values of the AC voltage setting (recalculated to a
peak value) and the DC voltage setting.
Frequency setting limit (the lower limit
≤
the upper limit)
All
models
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Upper
limit
Factory default
550.0 Hz
Setting Range
1.0 Hz (AC mode : 40 Hz) to 550.0 Hz
Lower
limit
Factory default
1.0 Hz (AC mode : 40 Hz)
Resolution 0.1
Hz
Note: In the AC mode, the setting range is 40.0 Hz to 550.0 Hz.
10.15 Remote Sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing input
terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be compensated.
When the remote sensing is on, the output voltage detection point corrected by the AGC or Autocal
function is switched to the sensing input terminal. When AGC or Autocal is off, only the detection
voltage used for measurement display is switched.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected.
AGC/Autocal
Measurement
voltage, power, power factor
Off On
On
Use the sensing input terminal voltage
Not active
Active
Off
Use the output terminal voltage
Not active (factory
default)
Active
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......