10. Specifications
PROGRAMMABLE AC POWER SOURCE
260
10.19 Simulation (Optional)
This allows you to simulate power line abnormalities, such as blackout, voltage rise, voltage drop,
abrupt phase change, and abrupt frequency change. It can be used only for AC, sine wave, and
AC+DC-INT. Note that it does not support IEC or other standard test. In the polyphase model,
polyphase system, and polyphase output of the Multi-phase model, only the balanced mode is
available.
All
models
Number of memories
5 (non-volatile)
Number of steps
6 (Initial, Normal 1, Trans 1, Abnormal, Trans 2, Normal 2)
Setting range of Step
Time
0.0010 s to 999.9999 s (0 s is available only for the Transition Step)
Parameter Output
range
(The above item is common within the Simulation)
AC voltage
Frequency
Waveform (sine wave only)
Start Phase (excluding the Transition Step)
Stop Phase (excluding the Transition Step)
Synchronous step output (2bit)
Trigger output
Repeat count (1 to 9999 times or infinite)
初期
定常1
移行1
異常
移行2
定常2
初期
トリガ設定例
ステップ同期1出力例
ステップ同期2出力例
Initial
Initial
Normal1
Normal2
Trans1
Trans2
Abnormal
Trigger setting example
Step sync 1 output example
Step sync 2 output example
Simulation control
Start
Stop
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......