10.23 General Function
DP Series
263
10.22.3 External Signal Input (Optional, only EXT and ADD)
This multiplies the input signal by the specified gain and outputs it. For ADD, the internal signal
source is added.
EXT: Output voltage (V) = External signal input (V) x Gain (V/V)
ADD: Output voltage (V) = External signal input (V) x Gain (V/V) + Internal signal source setting
(V)
Specification
Factory
default
100 V range: 0.0 to 220.0 times
100
Setting Range for gain
200 V range: 0.0 to 440.0 times
200
Setting resolution
0.1
Gain accuracy
±5 %
(DC, 45 Hz to 65 Hz, the gain is the factory default, the rated output
voltage, no load)
Input-output phase
In-phase
Input terminal
BNC connector (rear panel, unbalanced)
Also used as the external synchronization signal input
Input impedance
1 M
Ω
Input voltage range
±
2.2 V (A/D resolution 10 bit)
Nondestructive
maximum input
voltage
±10 V
Input frequency range
DC to 550 Hz (sine wave)
DC to 100 Hz (other than sine wave)
Note: There is no optional setting for the polyphase model. It cannot be used for the polyphase
system or polyphase output of the Multi-phase.
10.23 General Function
Description
Factory
default
Contrast
0 to 99
LCD
display setting Color
Blue tone or white tone
Blue tone
Beep
(key operation,
erroneous operation)
On or Off
Alarms on abnormal situation regardless of the setting
On
Key lock
On or Off
On: Only key lock-off and output-off are available
Off
Output relay control
On: The output relay is used to turn the output on/off
Off: The output relay is not used. High impedance to
turn the output off
On
Output setting
at power-on
On or Off
On: Output on after power-on
Off
Trigger output setting
Polarity: positive
or negative
Pulse width: 0.1 ms to 10 ms (resolution 0.1 ms)
Negative
10 ms
Time unit setting for
Sequence and Simulation
ms or s
s
Reset function
Resets the items stored in the System Setting Memory (excluding the
external interface setting) and the items that are to be reset at
power-on, to the factory default settings.
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......