3. Basic Operation
PROGRAMMABLE AC POWER SOURCE
58
3.4.9 Using the Measurement Function
The measured values are displayed in the measured value area (see
). The main measurement
functions of this product are shown in Table 3-11. For details of the display items of the measured
value, see
.
Table 3-11 Main Measurement Functions
Item Description
Unit
V Output
voltage
effective
value
The effective (RMS) value of the output
voltage on the output terminal.
Vrms
Vavg Output
voltage
average
value
The average DC value of the output
voltage on the output terminal.
V
Vmax
Output voltage maximum
peak value
The maximum peak value of the output
voltage on the output terminal.
Vpk
Vmin
Output voltage minimum
peak value
The minimum peak value of the output
voltage on the output terminal.
Vpk
I
Output RMS current
The effective (RMS) value of the output
voltage. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Arms
Iavg
Output current average
value
The average DC value of the output
current.
A
Imax Output
current
maximum
peak value
The maximum peak value of the output
current. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Apk
Imin
The minimum peak value
of the output current
The minimum peak value of the output
current. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Apk
Ipk-Hold Output current peak-hold
value
The hold value of Imax or Imin, either of
which the value is bigger. When the line
voltage display is L1-L2/L2-L3/L3-L1,
the phase current of the L1/L2/L3 phase
is displayed respectively.
Apk
P
Active power
The active power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
W
S
Apparent power
The apparent power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
VA
Q
Reactive power
The reactive power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
var
Summary of Contents for DP015S
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL NF Corporation ...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP Series INSTRUCTION MANUAL DA00027113 001 ...
Page 4: ......
Page 18: ......
Page 24: ...1 Outline PROGRAMMABLE AC POWER SOURCE 6 ...
Page 224: ...5 Description of Screen and Menu PROGRAMMABLE AC POWER SOURCE 206 ...
Page 238: ...7 Options PROGRAMMABLE AC POWER SOURCE 220 ...
Page 298: ......
Page 300: ......
Page 302: ......
Page 303: ......