11.6 AC Output
DP060LM/DP120LM/DP180LM
251
11.6 AC
Output
[V]=Vrms, [A]=Arms, unless otherwise noted.
Single-phase output
DP060LM DP120LM DP180LM
Mode Single-phase
two-wire
Floating output, the Lo terminal can be grounded.
Rated output voltage
100 V / 200 V
Voltage setting range
0.0 V to 160.0 V / 0.0 V to 320.0 V
0.0 Vp-p to 454.0 Vp-p / 0.0 Vp-p to 908.0 Vp-p (arbitrary wave)
Setting resolution
0.1 V
Voltage accuracy *1
(0.5 % of set + 0.6 V / 1.2 V)
Maximum current *2 60 A / 30 A
120 A / 60 A
180 A / 90 A
Maximum peak
current *3
Peak value (Apk) which is four times of the
maximum current
Peak value (Apk) which
is three times of the
maximum current
Short reverse power
flow *4*5
100 % or less of Maximum current (RMS)
(reverse power flow time
≦
20 ms, discontinuous, 40 °C or lower)
Power capacity
6 kVA
12 kVA
18 kVA
Load power
factor
0 to 1 (phase lead or phase lag, 45 Hz to 65 Hz, external power injection and
regeneration are not available.)
Frequency
setting range
40.00 Hz to 550.00 Hz (AC mode)
1.00 Hz to 550.00 Hz (ACDC mode)
Setting resolution
0.01 Hz
Frequency accuracy
0.01 % of set (23 °C
5 °C)
Frequency stability
*6
0.005 %
Voltage frequency
characteristic *7
1 %
Output waveform
sine wave, arbitrary wave (16 types), clipped sine wave (3 types)
Output on phase
setting range
0.0°
to 359.9° variable
Setting
resolution 0.1°
Output off phase
setting range
0.0°
to 359.9° variable (active/inactive selectable)
Setting
resolution 0.1°
DC offset
*8 Within
20 mV (typ. fine adjustment available)
*1
:
In the case of 10 V to 150 V/20 V to 300 V, sine wave, no load, 45 Hz to 65 Hz, DC voltage setting
0 V, 23 °C
5 °C.
*2
:
If the output voltage is higher than the rated value, this is limited (lowered) to satisfy the power
capacity. If there is the DC superimposition, the active current of AC+DC satisfies the maximum
current. In the case of 40 Hz or lower or 400 Hz or higher, and that the ambient temperature is 40°C
or higher, the maximum current may decrease.
*3
:
For the capacitor input type rectified load (crest factor=4 or 3), the rated output voltage, and 45 Hz to
65 Hz.
*4
:
In the case rated output voltage, 50 Hz or 60 Hz. If the output voltage is higher than the rated value,
this is limited to satisfy the power capacity. It may reduce short reverse power flow if ambient
temperature is 40°C or higher or repeated interval of reverse power flow is 15 minutes or less.
*5
:
External power injection or regeneration which is over short reverse power flow capacity is not
available.
*6
:
For 45 Hz to 65 Hz, the rated output voltage, no load or the resistance load for the maximum current,
and within the operating temperature.
Summary of Contents for DP060LM
Page 1: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...PROGRAMMABLE AC POWER SOURCE DP060LM DP120LM DP180LM INSTRUCTION MANUAL DA00059920 004...
Page 4: ......
Page 19: ...DP060LM DP120LM DP180LM 1 1 Outline 1 1 Overview 2 1 2 Series Lineup 2 1 3 Features 3...
Page 24: ......
Page 224: ......
Page 244: ......
Page 264: ......
Page 300: ...11 Specifications PROGRAMMABLE AC POWER SOURCE 282 Figure 11 4 DP180LM Type5L cabinet...