10.
Specifications
PROGRAMMABLE AC/DC POWER SOURCE
242
immediately after a change of the input power-supply voltage is not included.
*19: In the case that the output current is changed from 0% to 100% of the maximum current. For
output voltage 75 V to 150 V/150 V to 300 V, no load reference. However, if the output
voltage is higher than the rated value, the maximum current is limited to satisfy the power
capacity.
*20: For power input 200 V, no load, the rated output voltage, DC or 45 Hz to 65 Hz.
10.8 Distortion of Output Voltage Waveform
KP3000S
Distortion
(Phase voltage)
0.5% or lower
(40 Hz to 550 Hz, 50% or higher of the rated output voltage, the maximum
current or lower, AC and AC+DC modes, THD+N)
10.9 Power
Input
KP3000S
Voltage
100 V to 230 V
10% (250 V or lower) Overvoltage Category II
Frequency
50 Hz
2 Hz or 60 Hz
2 Hz
Phase Single-phase
Power factor *21
Power input 100 V: 0.95 or higher (typ.)
Power input 200 V: 0.90 or higher (typ.)
Efficiency
*21
77% or higher (typ. power input voltage 200 V)
KP3000S
Maximum power
consumption
4.5 kVA or lower
*21: In the case of AC- INT, the rated output voltage, the resistance load at the maximum current,
45 Hz to 65 Hz output.
10.10 Withstand Voltage and Insulation Resistance
Power input - Output/Chassis and Power input/Chassis - Output
KP3000S
Withstand voltage
AC 1500 V or DC 2130 V 1 minute
Insulation resistance
30 M
or higher (DC 500 V)
Summary of Contents for KP3000S
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Page 4: ......
Page 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Page 22: ......
Page 42: ......
Page 78: ......
Page 210: ......
Page 220: ......
Page 252: ......
Page 286: ......
Page 288: ......
Page 289: ......