10.
Specifications
PROGRAMMABLE AC/DC POWER SOURCE
252
10.20 Clipped Sine Wave
The peak clipped sine wave can be output, based on the crest factor (CF) setting or the percent
setting to the peak value.
KP3000S
Number of memories
3 (non-volatile)
CF
Variable range
1.10 to 1.41
Factory default
1.41
Setting resolution
0.01
Effective value correction
Yes
Clip ratio
Variable range
40.0% to 100.0%
Factory default
100.0%
Setting resolution
0.1%
Effective value correction
None
Note: The crest factor is represented as "voltage peak value/voltage effective value." It is 1.41 for
sine wave.
Note: When the clip ratio is specified, the peak is clipped by the voltage corresponding to the
specified % to the peak value of the setting voltage (100%).
Example) For the output voltage setting of 100 Vrms and the clip rate of 80%, the peak is
clipped at 113.1 Vpk.
Note: In the polyphase system, these are the settings for the phase voltage.
10.21 Arbitrary Wave
This uses the waveform data saved in the internal memory, which is transferred and recalled using
the external interface or USB memory.
KP3000S
Number of memories
16 (non-volatile)
Waveform length
4096 words
Amplitude resolution
16 bit
Summary of Contents for KP3000S
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Page 4: ......
Page 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Page 22: ......
Page 42: ......
Page 78: ......
Page 210: ......
Page 220: ......
Page 252: ......
Page 286: ......
Page 288: ......
Page 289: ......