4. Advanced Operation
PROGRAMMABLE AC/DC POWER SOURCE
70
Step
One sequence consists of a series of two or more steps. To use Sequence function, first set each step.
Step-control Parameters and intra-Step Parameters
There are the following two major types of parameters in Sequence function:
Step-control parameter
Parameter to control the flow between two or more steps. This defines duration of each step
and the next step to transition. Settings of step synchronous code and trigger output are
included in step-control parameter for convenience.
Intra-Step parameter
Parameter to define output state in each step. This defines basic parameters such as a
waveform, frequency, and how they are changed in the step.
Figure 4-1 Step-control Parameters and intra-Step Parameters
Step No. 0
Step No. 0 is assigned to the stand-by state before the sequence starts. If the terminal specification
of the step other than No. 0 is End, the execution transitions to step No. 0 and enters into the
stand-by state. Step Time and the Step Behavior setting of No. 0 specifies the action to transit to No.
0 at the end of a sequence.
Step number
Step-control parameter
Intra-Step parameter
Icon
Summary of Contents for KP3000S
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Page 4: ......
Page 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Page 22: ......
Page 42: ......
Page 78: ......
Page 210: ......
Page 220: ......
Page 252: ......
Page 286: ......
Page 288: ......
Page 289: ......