4. Advanced Operation
PROGRAMMABLE AC/DC POWER SOURCE
72
4.2.2 Parameters of Sequence Function
Table 4-2 shows the parameters of Sequence function. The common parameters are common to one
sequence. In Sequence function, signal source is fixed to INT. In each step, set the step-control
parameter and the intra-Step parameter. As for intra-Step parameter, the item and range that can be
set differ depending on the output range (100 V/200 V range) and output mode
(AC-INT/ACDC-INT/DC-INT) as in the case of continuous output function. This section describes
intra-Step parameters and step-control parameters separately.
Table 4-2 Parameters of Sequence Function
Common
parameter
Step-control parameter
Intra-Step parameter
Output range
AC/DC mode
Step Time
Specification of the Jump-to step
Jump count (1 to 9999, or infinite)
Specification of the branch step
Start phase
Stop Phase
Step termination
Synchronous step output (2bit)
Trigger output
Frequency
Waveform
Phase angle
AC phase voltage
DC voltage
Frequency (Freq), waveform (Wave), phase angle (Phase), AC phase voltage (ACV), DC
voltage (DCV)
The setting for the output of each step. In polyphase system, AC phase voltage and phase angle can
be set for each phase. In the phase angle, set the phase difference with the L1 phase. Phase angle can
also be set per step. The phase angle of L1 phase cannot be set, and is fixed as 0 deg.
Step Time (Step Time)
Set the output time of that step. Unit can be selected from [s] or [ms].
Jump-to step (Jump To)
When selected, the destination step of the jump at the end of the step can be specified. By specifying
the Jump count ([Jump Cnt]), you can set up a loop to repeat the same step. After jumping for the
times specified for one sequence, the action at the end of step follows the step termination setting.
If Jump-to is set to off, the action at the end of step follows the step termination setting.
--------
Notes
------------------------------------------------------------------------------------------------------------
When the step termination is set as [Hold], the hold status is enabled before the jump.
-----------------------------------------------------------------------------------------------------------------------------
Summary of Contents for KP3000S
Page 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Page 2: ......
Page 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Page 4: ......
Page 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Page 22: ......
Page 42: ......
Page 78: ......
Page 210: ......
Page 220: ......
Page 252: ......
Page 286: ......
Page 288: ......
Page 289: ......