3.2 Input and output connections
WF1943B
3-9
Trigger/sweep input (TRIG/SWEEP IN)
Signal characteristics:
The following types are produced during the burst oscillation mode.
Trigger: Select oscillation start for either
or
.
Gate: Select oscillation at either high or low level.
Triggered gate: Oscillation start/stop at
or
. Select either.
Minimum pulse width is 50 ns.
In sweep oscillation mode, the sweep starts at
or
. Select either.
Minimum pulse width is 200 ns.
Input voltage :
High level
≥
+3.9 V
Low
level
≤
+1.6 V
Input voltage range :
−
0.5 to +5.5 V
Input impedance :
Pull up to +5 V at approx. 10 k
Ω
Ground :
Connected to chassis ground
Do not apply a signal exceeding the above input voltage range. The unit can be damaged.
• Drive circuit examples
Connect the trigger and sweep input drive signals to TTL or C-MOS logic IC outputs.
Since the input circuit is provided with a built-in pullup resistor, an open collector output drive can also be
used. However, contact chatter from a mechanical switch or relay can prevent normal operation. Also,
chattering will prevent normal operation when the oscillation mode is triggered gate.
Avoid using a logic IC circuit having a power supply voltage higher than +5 V such as example (c) for the
WF1943B input
! CAUTION
+5 V
+12 V
~
(a) TTL logic output (b) Open collector output (c) High voltage logic output
TRIG/SWEEP IN
+5 V
10 k
Ω
100
Ω
D
Summary of Contents for WF1943B
Page 2: ......
Page 3: ...DA00012503 002 WF1943B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 83: ...4 2 Sweep WF1943B 4 35 Sweep value and SYNC OUT Approximately 100 to 125 s...
Page 150: ......
Page 152: ......
Page 154: ......
Page 155: ......