WF1945B
7-3
7.2 Operation checks
Preparatory checks
Check the following before inspection.
z
Power source voltage is within specified range.
z
Ambient temperature is within 5 to 35
°
C.
z
Ambient relative humidity is within 10 to 85 %RH (absolute humidity: 1 to 25 g/m
3
)
z
Condensation is absent.
Function checks
z
Power on
Confirm absence of error message at power on.
If an error message appears,
“6. Troubleshooting”, cf
.
If an abnormal indication appears at power on, switch power off and wait at least 5 seconds, then
again switch power on.
z
Main function checks
To avoid setting error, initialize the settings (
SYSTEM
→
PRESET).
Connect FUNCTION OUT to an oscilloscope with 50
Ω
coaxial cable to monitor the output.
Operate to change the following settings several times and confirm normal functions. Test both
keypad and MODIFY dial for settings such as frequency.
・
Frequency (
ENTRY
→
FREQ
)
・
Amplitude (
ENTRY
→
AMPTD
)
・
DC offset (
ENTRY
→
OFFSET
)
・
Waveform (
FUNCTION
→
,
,
Duty fixed 50 %
,
,
)
・
Duty (
FUNCTION
→
Variable duty
,
ENTRY
→
DUTY
)
・
Output on/off (
OUTPUT
ON/OFF
)
Summary of Contents for WF1945B
Page 2: ......
Page 3: ...DA00012505 002 WF1945B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 30: ......
Page 62: ......
Page 100: ...4 2 Sweep 4 38 WF1945B Sweep value and Z MARKER SYNC X DRIVE outputs is approx 100 to 125 s...
Page 101: ...4 2 Sweep WF1945B 4 39...
Page 128: ......
Page 158: ......
Page 182: ......
Page 184: ......
Page 186: ......
Page 187: ......