4.3.4 Output Mode Setting
Please press the button below
Mode
on the screen to enter output mode setting.
The default is normal (NORM) mode. In normal mode, channel 1, channel 2, and
channel 3 output separately and are set individually. The other three optional output
modes are parallel (PARALL), series (SERIES), and trace (TRACE) modes. Users can
switch to the corresponding mode by pressing the corresponding button below the
different modes, as shown in the figure below.
Figure 12 Output Mode Setting
4.3.4.1 Parallel Mode
Please press the button below
PARALL
on the screen and switch to parallel mode.
Figure 13 Parallel Mode
In
parallel
mode,
the
power
supply
can
output
0-6A
for
N3411E/N3413E/N3411/N3411P/N3413/N3413P
and
0-10A
for
N3412E/N3412/N3412P.
The
power
supply
can
output
0-32V
for