Hunan Next Generation Instrumental T&C Tech. Co., Ltd.
NGI
l
NGI
N3410 Series
Triple-channel Programmable DC Power Supply
19
Figure 17 Trace Mode
In trace mode, channel 2 follows channel 1 to output the same value of negative
voltage and negative current. Channel 2 trace voltage output can be set to -32V, and
current output can be set to -3A, following channel 1. The trace connection method
is shown in the figure below.
-
-
-
CH1
CH2
CH3
Load
Figure 18 Trace Mode Wiring Diagram
Steps to set voltage and current in trace mode:
1. Press
or
Enter
on voltage/current setting area.
2. Input digit by pressing numeric buttons or pressing
to move the cursor
and rotate
to adjust the value.
3. Press
or
Enter
to confirm the setting.
After completing setting, please press
C H 1
. ON icon will show at channel 1 and
channel 2. Channel 2 follows channel 1 to output the same voltage and current in