N3410 Series
Triple-channel Programmable DC Power Supply
14
5. Press
or
Enter
to confirm the setting.
Figure 9 Voltage/Current Setting
4.3.3 Protect Setting
The OVP/OCP setting range of N3410 series is shown below. By default, OVP/OCP is
set to the maximum value.
Table 7
Model
N3411
N3412
N3413
OVP Setting Range (CH1/CH2)
3~34.12V
3~34.12V
3~63.5V
OCP Setting Range (CH1/CH2)
1~3.25A
1~5.35A
1~3.25A
OVP Setting Range (CH3)
3~6.8V
OCP Setting Range (CH3)
1~3.25A
Steps to set protection parameters:
1. Press button below
Protect
on the screen.
2. Press
or
Enter
on OVP/OCP setting area.
3. Input digit by pressing numeric buttons or pressing
to move the cursor
and rotate
to adjust the value.
4. Press
or
Enter
to confirm the setting.
5. Press button below
Channel
on the screen to switch to other channels.
6. Press
<
to return to previous page.