background image

14

|

ni.com

|

NI PXIe-5624R Calibration Procedure

5.

Configure the NI 5624R with the following settings:

DDC mode

Center frequency: The first frequency listed in the 

NI 5624R Center Frequency (MHz) 

column in Table 3.

I/Q rate: 500 MHz

Dither: On (High)

6.

Configure the signal generator with the following settings: 

Frequency: The first frequency listed in the 

Generator Start Frequency (MHz) 

column 

in Table 3

Power level: 4 dBm

7.

Configure the power sensor to correct for the current frequency of the signal generator 
using the power sensor frequency correction function.

8.

Use the power sensor to measure the power. Record the result as 

measured input power

.

Note

Power sensor readings are in dBm, not watts.

9.

Calculate the 

corrected input power

 using the following equation:

corrected input power 

measured input power

 + 

splitter balance

Tip

Find the splitter balance by using the appropriate data points from the 

Characterizing Power Splitter Balance

 section.

10. Perform an acquisition and measure the tone present at the frequency generated by the 

signal generator. 

11. Calculate the 

absolute amplitude accuracy

 using the following equation:

absolute amplitude accuracy = NI 5624R input power – corrected input power 

Table 3.  

Configured Center Frequencies

NI 5624R Center 

Frequency (MHz)

Generator Start 

Frequency (MHz)

Generator Stop 

Frequency (MHz)

Step Size (MHz)

225

25

425

5

500

300

700

5

800

600

995

5

1200

1005

1400

5

1500

1300

1700

5

1800

1600

1975

5

Summary of Contents for 5624R

Page 1: ...PGA converts data from the ADC into quadrature I Q data This DSP shifts the frequency of decimates equalizes and calibrates the data before storing it into DRAM Use Interactive Spectrum Acquisition Ho...

Page 2: ...oftware from ni com downloads Documentation You might find the following documentation helpful as you perform the calibration procedure NI PXIe 5624R Getting Started Guide NI IF Digitizers Help NI PXI...

Page 3: ...for this procedure Verifying absolute amplitude accuracy Adjusting internal frequency reference B22 option needed for this procedure Adjusting absolute amplitude accuracy Frequency range 4 MHz to 2 00...

Page 4: ...adapter Huber Suhner 31_3 SMA 50 1 1 _UE or Fairview Microwave SM4236 Test system characterization Frequency range DC to 2 GHz Impedance 50 Return loss 23 dB 3 5 mm m to 3 5 mm m adapter Huber Suhner...

Page 5: ...each verification procedure insert a delay between configuring all devices and acquiring the measurement Adjust this delay depending on the instruments used but make sure it is at least 5 000 ms for t...

Page 6: ...e verification and adjustment procedures use test points from 5 MHz to 2 005 GHz in 5 MHz steps for the splitter balance characterization Frequencies in which harmonics alias on the fundamental freque...

Page 7: ...rrection function 7 Use the power sensor to measure the power at the frequency from step 6 8 Repeat steps 5 through 7 by updating the signal generator frequency from 5 MHz to 2 005 GHz in 5 MHz steps...

Page 8: ...e signal generator frequency from 5 MHz to 2 005 GHz in 5 MHz steps Record the resulting measurements as splitter output 1 power Each frequency should have a corresponding value 16 Calculate the split...

Page 9: ...ate self calibration using one of the following methods Click Self Calibrate in the upper right corner of MAX Right click the name of the device in the MAX configuration tree and select Self Calibrate...

Page 10: ...ference using a signal generator Use digitizer mode of the NI 5624R Use Interactive IF Acquisition Without DSP Host vi in the Multirecord Acquisition sample project as a starting point 1 Connect the s...

Page 11: ...gle Tone Information vi 5 Calculate the deviation using the following equation 6 Compare the value determined in step 5 to the verification test limits in the following table where initial accuracy 20...

Page 12: ...e across 400 MHz of bandwidth for each center frequency is measured This procedure requires the test setup and data collected in the Characterizing Power Splitter Balance section Characterize the powe...

Page 13: ...front panel connector The following figure illustrates the complete hardware setup Figure 4 Absolute Amplitude Accuracy Verification Cabling Diagram 1 Power Sensor 2 SMA m to N f Adapter 3 Power Spli...

Page 14: ...sult as measured input power Note Power sensor readings are in dBm not watts 9 Calculate the corrected input power using the following equation corrected input power measured input power splitter bala...

Page 15: ...R flash memory Successfully complete all required verifications or successfully complete reverification following adjustment prior to updating the calibration date and time on the NI 5624R Complete th...

Page 16: ...panel connector to the NI 5624R IF IN front panel connector using an SMA m to SMA m cable The following figure illustrates the hardware setup Figure 5 Internal Frequency Reference Adjustment Cabling...

Page 17: ...he power splitter balance characterization data for the correct frequency as the following frequencies are excluded from the adjustment procedure due to harmonics aliasing back on to the fundamental a...

Page 18: ...that the NI 5624R and the signal generator share a common reference The following steps assume you have connected the signal generator reference frequency REF OUT to the NI 5624R REF IN connector 6 C...

Page 19: ...plitter output 1 of the power splitter using the power sensor Record this value as the measured input power 12 Calculate the corrected input power using the following equation corrected input power me...

Page 20: ...NI Refer to the Worldwide Support and Services section for information about support resources or service requests Worldwide Support and Services The National Instruments website is your complete reso...

Reviews: