Initialization software uses the configuration information specific to each adjacent peripheral
module to evaluate local bus compatibility.
PXI Trigger Bus
All slots share eight PXI trigger lines. You can use these trigger lines in a variety of ways. For
example, you can use triggers to synchronize the operation of several different PXI peripheral
modules. Modules can pass triggers to one another on the lines, allowing precisely timed
responses to asynchronous external events the system is monitoring or controlling.
Dynamic routing of triggers (automatic line assignments) is supported through certain NI
drivers like NI-DAQmx.
Note
Although any trigger line may be routed in either direction, it cannot be
routed in more than one direction at a time.
System Reference Clock
The PXIe-1090 chassis supplies PXI_CLK10, PXIe_CLK100, and PXIe_SYNC100 to every
peripheral slot with an independent driver for each signal.
An independent buffer (having a source impedance matched to the backplane and a skew of
less than 250 ps between slots) drives PXI_CLK10 to each peripheral slot. You can use this
common reference clock signal to synchronize multiple modules in a measurement or control
system.
An independent buffer drives PXIe_CLK100 to each peripheral slot. These clocks are matched
in skew to less than 100 ps. The differential pair must be terminated on the peripheral with
LVPECL termination for the buffer to drive PXIe_CLK100 so that when there is no peripheral
or a peripheral that does not connect to PXIe_CLK100, there is no clock being driven on the
pair to that slot.
An independent buffer drives PXIe_SYNC100 to each peripheral slot. The differential pair
must be terminated on the peripheral with LVPECL termination for the buffer to drive
PXIe_SYNC100 so that when there is no peripheral or a peripheral that does not connect to
PXIe_SYNC100, there is no SYNC100 signal being driven on the pair to that slot.
PXI_CLK10, PXIe_CLK100, and PXIe_SYNC100 have the default timing relationship
described in the following figure.
Figure 5. System Reference Clock Default Behavior
PXIe_CLK100
PXI_CLK10
PXIe_SYNC100
0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9
PXIe-1090 User Guide
|
© National Instruments Corporation
|
7