NI 9949 Accessory User Guide

2

ni.com

NI 9949 Dimensions

The NI 9949 has outer dimensions of 65.6 × 21.3 × 31.5 mm 
(2.58 × 0.84 × 1.24 in.), as shown in Figures 2 and 3.

Figure 2.  

NI 9949 Dimensions

Figure 3.  

NI 9949 Dimensions and Hole Clearance

Connecting the NI 9949

Use an RJ-50 cable to connect the NI 9949 to devices with RJ-50 
receptacles, such as the NI 9237 with RJ-50 and the SCXI-1314T. The 
NI 9949 routes each pin of the RJ-50 receptacle to the corresponding screw 
terminal, where 1–10 represent the pin and terminal numbers and S 
represents the shield of the RJ-50 connector. For example, if you connect 
the NI 9949 to the NI 9237, pin 1 of the NI 9237 RJ-50 receptacle connects 
to terminal 1 of the NI 9949 screw-terminal connector. Refer to Figure 4 
for a diagram of the RJ-50 and screw-terminal connectors on the NI 9949. 
Refer to the device manual for information about device pin assignments.

Note

National Instruments does not recommend using an RJ-45 cable with the NI 9949 

because it can physically damage pins 1 and 10 on the device, and thus permanently 
disable shunt calibration no matter what connector you use.

65.6 mm (2.58 in.)

21.3 mm
(0.84 in.)

31.5 mm
(1.24 in.)

Holes for
Tie Strap

Ø 4.6 mm (0.18 in.)

Ø 8.9 mm (0.35 in.)