You can download all required software from
ni.com/downloads
.
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•
NI Digital Multimeters Help
•
NI PXIe-4080/4081/4082 Getting Started Guide
•
NI PXIe-4081 Specifications
Visit
ni.com/manuals
for the latest versions of these documents.
Password
The default password for password-protected operations is
NI
.
Calibration Interval
Recommended calibration interval
2 years
Test Equipment
The following table lists the equipment required for calibrating your device. If you do not have
the recommended instruments, use these specifications to select a substitute calibration
standard.
Equipment
Recommended
Models
Where Used
Minimum Requirements
Multifunction
calibrator
Fluke 5720A or
Fluke 5730A
Voltage, Current,
Resistance
Maintained at 90 day
specifications
Three sets of low
thermal
electromotive force
(EMF) copper
cables
Fluke 5440
Voltage and
>1 mA Current
(1 set),
Resistance (2
sets), ≤1 mA
Current (3 sets)
Shielded twisted pair copper
cables with copper or gold-
plated copper banana plugs
Banana shorting bar Pomona 5145
DCV, 2-Wire
Resistance
≤ 40 mΩ resistance
2
|
ni.com
|
PXIe-4081 Calibration Procedure