DMM Settings
Calibrator Settings
Function
Range
Amplitude
Frequency
AC Volts
50 V
50 V
1 kHz
5 kHz
20 kHz
50 kHz
100 kHz
200 kHz
300 kHz
10 V
500 kHz
AC Volts
700 V
100 V
1 kHz
5 kHz
20 kHz
50 kHz
100 kHz
200 kHz
50 V
300 kHz
10 V
500 kHz
5.
Reset the calibrator.
6.
Call the niDMM_Cal Adjust Misc VI or the
niDMM_CalAdjustMisc
function with
Type
set to
NIDMM_EXTCAL_MISCCAL_VOLTAGE_AC_FILTER
.
7.
Self-calibrate the PXIe-4081.
You have completed adjusting the AC voltage modes of the PXIe-4081. Proceed to one of the
following sections:
•
Adjusting Current
If you are not performing additional adjustments, proceed to the
Completing the Adjustment
Procedures
section to commit the calibration constants.
Adjusting Current
Note
You must adjust DC voltage before you adjust DC current.
PXIe-4081 Calibration Procedure
|
© National Instruments
|
25