5.
Close the NI-DCPower session.
Voltage Output and Measurement
Connecting and Configuring Equipment for Voltage Verification
1.
Make the necessary connections for this procedure, as shown in the following figure:
Figure 2. Voltage Verification or Adjustment Connection Diagram
DMM
Voltage
Mode
+
–
NI-DCPower Device
Sense LO
Sense HI
HI
LO
2.
Set the
niDCPower Output Function
property or
NIDCPOWER_OUTPUT_FUNCTION
attribute to DC Voltage for the NI 4141.
Verifying Voltage Output and Measurement
Compare a set of voltages measured by the external DMM to both a set of voltage test points
requested by the NI 4141 and to the measured voltages reported by the NI 4141.
Refer to the following table as you complete the following steps.
6
|
ni.com
|
NI PXIe-4141 Calibration Procedure