background image

Isolation

Caution

 

Isolation voltage ratings apply to the voltage measured between any

channel pin and the chassis ground. When operating channels in series or floating on
top of external voltage references, ensure that no terminal exceeds this rating.

Attention

 

Les tensions nominales d'isolation s'appliquent à la tension mesurée

entre n'importe quelle broche de voie et la masse du châssis. Lors de l'utilisation de
voies en série ou flottantes en plus des références de tension externes, assurez-vous
qu'aucun terminal ne dépasse cette valeur nominale.

Caution

 

Do not connect the TB-414X to signals or use for measurements within

Measurement Categories II, III, or IV.

Attention

 

Ne connectez pas le TB-414X à des signaux et ne l'utilisez pas pour

effectuer des mesures dans les catégories de mesure II, III ou IV.

Measurement Category I is for measurements performed on circuits not directly connected to
the electrical distribution system referred to as 

MAINS

 voltage. MAINS is a hazardous live

electrical supply system that powers equipment. This category is for measurements of voltages
from specially protected secondary circuits. Such voltage measurements include signal levels,
special equipment, limited-energy parts of equipment, circuits powered by regulated low-
voltage sources, and electronics.

Note

 

Measurement Categories CAT I and CAT O are equivalent. These test and

measurement circuits are for other circuits not intended for direct connection to the
MAINS building installations of Measurement Categories CAT II, CAT III, or
CAT IV.

Isolation voltage, channel to earth ground

5

60 V DC, CAT I

Withstand voltage

800 V

pk

Screw Terminal Wiring

Screw terminal wire gauge

24 to 18 AWG

Length of bare wire (insulation stripped)
from one end of the wire

5 mm to 6 mm (0.197 in. to 0.236 in.)

Screw terminal torque

0.5 Nm (4 lb · in.)

Spacing between screw terminals

3.5 mm (0.14 in.)

5

While TB-414X channels are isolated from each other, this isolation is considered functional
isolation for the purpose of allowing the separation of return currents and is not intended to operate
continuously under significant voltages or provide safety isolation. The TB-414X is intended for
use with PXIe-414

x

 SMUs which are isolated from earth ground, but share a common LO for all

channels (bank isolation).

TB-414X User Guide 

  | 

© National Instruments Corporation

  | 

9

Summary of Contents for TB-414X

Page 1: ...nstalling and configuring your system Caution Observe all instructions and cautions in the user documentation Using the product in a manner not specified can damage the product and compromise the buil...

Page 2: ...when you see this icon printed on the model Cautionary statements are localized into French for compliance with Canadian requirements Getting Started with the TB 414X Kit Contents and Other Equipment...

Page 3: ...ze negative effects from wiring insulation materials and contaminants The SHDB25F DB25F LL cable accessory has been designed with these considerations in mind and utilizes the TB 414X allowing for one...

Page 4: ...the stripped end of the wires into the appropriate terminals and tighten the screws of the screw terminals to a torque of 0 5 Nm 4 lb in Refer to the following pinout signal information and PXIe 414x...

Page 5: ...H 1 C H 2 Table 1 TB 414X Signals and DB25 Pin Map Screw Terminal DB25 Connector Pin LO Connector CH 0 LO 16 CH 0 LO S 3 CH 1 LO 19 CH 1 LO S 6 CH 2 LO 22 CH 2 LO S 9 CH 3 LO 25 CH 3 LO S 12 CH S GND...

Page 6: ...he strain relief upward If you are connecting only a few wires assemble the strain relief downward for greater retention Figure 3 TB 414X Strain Relief 1 2 1 Strain Relief Up 2 Strain Relief Down b In...

Page 7: ...ations These specifications apply to the TB 414X for use with the PXIe 414x The PXIe 414x is a 4 Channel SMU Refer to the module specifications at ni com manuals for more information Definitions Warra...

Page 8: ...nel at the DUT end of the cable must ensure that LO path wiring resistances are matched within 20 to prevent the LO terminal of one channel from carrying a significant portion of another channel s LO...

Page 9: ...ements of voltages from specially protected secondary circuits Such voltage measurements include signal levels special equipment limited energy parts of equipment circuits powered by regulated low vol...

Page 10: ...ronmental Guidelines Notice This model is intended for use in indoor applications only Environmental Characteristics Temperature and Humidity Temperature Operating 0 C to 55 C6 Storage 40 C to 71 C 6...

Page 11: ...NI customers For additional environmental information refer to the Commitment to the Environment web page at ni com environment This page contains the environmental regulations and directives with whi...

Page 12: ...RMA by NI does not constitute export authorization The user must comply with all applicable export laws prior to exporting or re exporting this product See ni com legal export compliance for more inf...

Page 13: ...f help such as tutorials and examples Visit ni com services to learn about NI service offerings such as calibration options repair and replacement Visit ni com register to register your NI product Pro...

Page 14: ...LAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how...

Reviews: