background image

THE

Clifton

RANGE

DuoBaths
NE2D Series

Always maintain adequate water/oil levels.
Periodically check there is sufficient water/oil in the bath at all times.

Summary of Contents for Clifton RANGE NE2D Series

Page 1: ...THE CliftonRANGE DuoBaths NE2D Series Always maintain adequate water oil levels Periodically check there is sufficient water oil in the bath at all times...

Page 2: ...n not in use Explanation of Temperature control tems 9 Cleaning 10 Cleaning external painted surfaces featuring Anti bacterial finish Cleaning the stainless steel tank Descaling the stainless steel ta...

Page 3: ...of the over or under temperature alarm being illuminated do not touch the liquid as it may be very hot or cold 12 Recommend using a Lid above 60 C Take care when lifting the lid steam and hot vapours...

Page 4: ...ly or plug connection differs refer to local regulations or qualified electrician Liquid Level Minimum liquid level must cover the top of the false base by 40mm Maximum liquid level must not exceed th...

Page 5: ...each tank Indicators heating temperature alarm timer and set temperature each tank Keyboard function up and down arrow adjustment run buttons Drain outlet for accessory tap 22 litre size tanks Mains p...

Page 6: ...Press twice t is displayed timer setting DOWN ARROW Used to decrease a value Hold continously to scroll When pressed for more than 1 5 seconds SP1 is displayed UP ARROW Used to increase a value Hold...

Page 7: ...eatedly press FUNCTION button until t appears on the display It will then alternate between showing t and time displayed as hh mm 2 Press either up or down arrow to select desired number of hours minu...

Page 8: ...e set point When in alarm condition the over and under temperature alarm indicator illuminates and actual bath temperature is shown All heating is switched off Once water temperature has fallen below...

Page 9: ...ath by convection and conduction and there are heat losses from the surface of the liquid which can cause temperature losses These losses and heat distribu tion produce small fluctuations in temperatu...

Page 10: ...is maintained over its lifetime as the anti bacterial agent is integral within the paint In a laboratory environment it makes this one less source of contamination contributing to essential clean wor...

Page 11: ...ore detailed information re lating to how Virkon should be used with access to test reports www relyon dupont com Is Virkon solution corrosive Virkon solution requires only 10 minutes contact time to...

Page 12: ...rical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner...

Page 13: ...The Clifton Range 3 Year Warranty 13 High Quality Products at Affordable Prices...

Page 14: ...The Clifton Range 3 Year Warranty 14 High Quality Products at Affordable Prices...

Page 15: ...t tube rack into the bath on top of either the stainless steel perforated shelf or the raised shelf Note 4 Litre 1 Rack 8 Litre 2 Racks 14 Litre 4 Racks 22 and 28 Litre 6 Racks Heat Transfer Fluid Use...

Page 16: ...Spares Service Diagram The Clifton Range 3 Year Warranty 16 High Quality Products at Affordable Prices...

Page 17: ...The Clifton Range 3 Year Warranty 17 High Quality Products at Affordable Prices...

Page 18: ...The Clifton Range 3 Year Warranty 18 High Quality Products at Affordable Prices...

Page 19: ...acturers of laboratory medical and clinical equipment Oldmixon Crescent Weston super Mare North Somerset BS24 9BL United Kingdom Tel 44 0 1934 626691 Fax 44 0 1934 630300 Email clifton nickel electro...

Page 20: ...BX0835 www nickel electro co uk...

Reviews: