background image

About this Manual

This user Manual contains instructions which must be followed in order that the product is operated
correctly.

Please ensure that each chamber has fluid in it before switching on the product.

General Notes

Please observe the following safety precautions:
1.

Fill the tank with water prior to connection to power supply.

2.

Connect only to a power supply with the corresponding voltage to that specified on the rating
label positioned on the rear of the unit.

3.

Ensure the power supply has a safety earth (ground) terminal.

4.

Ensure the mains switch and power supply connector are accessible during use.

5.

The mains supply cord fitted to this products is a heat resistant type and should be replaced
by an equivalent type.

6.

Do not block ventilation slots during use and follow installation instructions.

7.

Always follow good laboratory practice by ensuring substances being heated present no risk of
a hazard (explosion, implosion or release of toxic or flammable gases) or that these have been
addressed. When heating substances where liberation of gases occurs suitable extraction
should be used.

8.

Use only liquids specified in this Instruction Manual within their specified temperature range.
If the alarm lamp is illuminated the liquid temperature may be above its recommended
maximum.

9.

Use caution when topping up or draining the tank as the liquid in the tank may be very hot or
cold.

10.

Drain before moving the bath. Allow the liquid to cool to 40°C before draining.

11.

In the event of the over or under temperature alarm being illuminated do not touch the liquid
as it may be very hot or cold.

12.

Recommend using a Lid above 60°C. Take care when lifting the lid - steam and hot vapours
can cause scalding.

13.

Always use the display or a thermometer to check the temperature - do not touch liquid.

14.

If this product is not used in accordance with these instructions, then basic safety protection
afforded by the water bath may be affected.

15.

Before using any cleaning or decontamination method except those recommended, check with
your distributor that the proposed method will not damage the equipment.

Amendments

Issue 1

April

1999

Initial issue instruction book.

Issue 2

September 2001

Revised, NE1 discon layout update

Issue 3

August

2002

Update content

Issue 4

March

2004

Product redesign

Issue 5

October

2005

Note: both tanks filled if only 1 is used.

Issue 6

July

2006

OPT function, turn tank OFF when not in use.

Issue 7

October

2006

Disinfection/sterilisation and temp. terms.

Issue 8

June

2010

2010 model K30.

Symbols

HOT SURFACES

Paragraphs marked by this symbol indicate that a potential hazard to your personal
safety exists from heated surfaces or other appendages on the outside or inside of
the equipment.

CAUTION

This icon accompanies text and/or other international symbols dealing with poten-
tial damage to equipment. When present, it indicates that there is a potential
danger of equipment damage may occur if information stated within the "CAUTION"
paragraph is not adhered to or procedures are executed incorrectly.

The Clifton Range - 3 Year Warranty

3

High Quality Products at Affordable Prices

PROTECTIVE EARTH OR GROUND TERMINAL

Protective earth conductor terminal.

Summary of Contents for Clifton RANGE NE2D Series

Page 1: ...THE CliftonRANGE DuoBaths NE2D Series Always maintain adequate water oil levels Periodically check there is sufficient water oil in the bath at all times...

Page 2: ...n not in use Explanation of Temperature control tems 9 Cleaning 10 Cleaning external painted surfaces featuring Anti bacterial finish Cleaning the stainless steel tank Descaling the stainless steel ta...

Page 3: ...of the over or under temperature alarm being illuminated do not touch the liquid as it may be very hot or cold 12 Recommend using a Lid above 60 C Take care when lifting the lid steam and hot vapours...

Page 4: ...ly or plug connection differs refer to local regulations or qualified electrician Liquid Level Minimum liquid level must cover the top of the false base by 40mm Maximum liquid level must not exceed th...

Page 5: ...each tank Indicators heating temperature alarm timer and set temperature each tank Keyboard function up and down arrow adjustment run buttons Drain outlet for accessory tap 22 litre size tanks Mains p...

Page 6: ...Press twice t is displayed timer setting DOWN ARROW Used to decrease a value Hold continously to scroll When pressed for more than 1 5 seconds SP1 is displayed UP ARROW Used to increase a value Hold...

Page 7: ...eatedly press FUNCTION button until t appears on the display It will then alternate between showing t and time displayed as hh mm 2 Press either up or down arrow to select desired number of hours minu...

Page 8: ...e set point When in alarm condition the over and under temperature alarm indicator illuminates and actual bath temperature is shown All heating is switched off Once water temperature has fallen below...

Page 9: ...ath by convection and conduction and there are heat losses from the surface of the liquid which can cause temperature losses These losses and heat distribu tion produce small fluctuations in temperatu...

Page 10: ...is maintained over its lifetime as the anti bacterial agent is integral within the paint In a laboratory environment it makes this one less source of contamination contributing to essential clean wor...

Page 11: ...ore detailed information re lating to how Virkon should be used with access to test reports www relyon dupont com Is Virkon solution corrosive Virkon solution requires only 10 minutes contact time to...

Page 12: ...rical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner...

Page 13: ...The Clifton Range 3 Year Warranty 13 High Quality Products at Affordable Prices...

Page 14: ...The Clifton Range 3 Year Warranty 14 High Quality Products at Affordable Prices...

Page 15: ...t tube rack into the bath on top of either the stainless steel perforated shelf or the raised shelf Note 4 Litre 1 Rack 8 Litre 2 Racks 14 Litre 4 Racks 22 and 28 Litre 6 Racks Heat Transfer Fluid Use...

Page 16: ...Spares Service Diagram The Clifton Range 3 Year Warranty 16 High Quality Products at Affordable Prices...

Page 17: ...The Clifton Range 3 Year Warranty 17 High Quality Products at Affordable Prices...

Page 18: ...The Clifton Range 3 Year Warranty 18 High Quality Products at Affordable Prices...

Page 19: ...acturers of laboratory medical and clinical equipment Oldmixon Crescent Weston super Mare North Somerset BS24 9BL United Kingdom Tel 44 0 1934 626691 Fax 44 0 1934 630300 Email clifton nickel electro...

Page 20: ...BX0835 www nickel electro co uk...