Metal top units:

The metal top plate should be cleaned using a damp cloth and a mild detergent solution.

Cleaning is made easier if spillages are attended to promptly. In any case, spillages of acids 
and alkalis MUST be removed immediately as these chemicals can attack and damage the 
the surface of the aluminium alloy. Ensure that the appropriate safety precautions are 
observed.

Preparation of Media

Take particular care when heating liquids having a high viscosity. Viscous liquids can act as thermal 
insulators and can cause thermal breakage of the glassware. This is very important with media 
solutions as the viscosity will usually increase as the temperature rises.

❖ 

Check that the stirring action is sufficient to agitate the whole of the liquid. Unstirred areas in the 
liquid can result in uneven heat transfer and “hot spots” in the glassware.  This can induce thermal 
stress and so cause failure.

❖ 

Check the stirring action regularly to ensure that it remains adequate as the viscosity of the solution 
increases.

❖ 

Always use the largest magnetic follower possible and, if necessary, use a mechanical overhead 
stirrer.

❖ 

Do not use glass vessels with thick walls, e.g. Pyrex Heavy Duty Ware or standard beakers and flasks 
having capacities of 5 litres or greater.

❖ 

NEVER heat glass bottles on a hotplate.

❖ 

Ensure that the heat is built up slowly to avoid localised overheating.

❖ 

Ensure the glassware is completely free from scratches or  other defects. 

❖ 

Place the hotplate in a tray large enough to contain the liquid in the event of glassware failure.

❖ 

Wear the appropriate safety clothing - e.g. gloves, goggles, protective apron etc.

Accessory 

A rod is available to support apparatus used with the instrument. The instrument is equipped with a 
fitting on the rear to accept the support rod (see figure 2). To fit the rod to your instrument, first 
isolate from the mains supply and allow to cool. Then simply screw the threaded end of the support 
rod into the fitting on the rear of the instrument.

Servicing and Repair

 

This product range does not require any routine servicing.

Note: There are no internal user replaceable parts.

In the event of product failure it is recommended that any repair is only undertaken by suitably 
qualified personnel. For advice or to receive a service manual please contact the Service Department 
of Nickel-Electro Ltd.
Only spare parts supplied by Nickel-Electro Ltd or its agent should be used. Fitting of non-approved 
parts may affect the performance of the safety features of the instrument and void warranty.

Note: The magnetic stirrer drive utilises strong magnets.

If in doubt, please contact the Service Department of;
Nickel-Electro Ltd, Oldmixon Crescent, Weston-Super-Mare, North Somerset BS24 9BL
Tel: +44 (0)1934 626691  Fax: +44 (0)1934 630300
Email: clifton@nickel-electro.co.uk   www.nickel-electro.co.uk

Summary of Contents for Clifton Series

Page 1: ...rs THE CliftonRANGE Round and Rectangular Boiling Baths with Energy Regulator 1000ER and 1100ER Series PDF created with pdfFactory Pro trial version www software partners co uk MSH 1 MSH 1D MSH 3 CH 1...

Page 2: ...ht Heater on light Stirrer on light Mains on light Hot warning light IEC socket External probe socket Retort rod fitting Digital display Stirrer control Heater control Digital models Figure 3 Front vi...

Page 3: ...odel Wattage CH 1E 550W MSH 1 750W CHS 1E 550W MSH 1D 750W CHS 2E 1200W MSH 3 3 x 700W There is an IEC socket at the rear of the instrument for connection to the mains supply see figure 2 The unit is...

Page 4: ...se the knob Repeat to turn off the heating function When switched on the display will show actual plate temperature at the bottom of the screen and the set temperature directly above The heating icon...

Page 5: ...equately 2 Keep the unit out of strong drafts 3 Allow the temperature adequate time to stabilise Safety Advice Do not use the instrument to heat flammable liquids Never lift or carry the instrument un...

Page 6: ...at glass bottles on a hotplate Ensure that the heat is built up slowly to avoid localised overheating Ensure the glassware is completely free from scratches or other defects Place the hotplate in a tr...

Page 7: ...under normal laboratory conditions for a period of three 3 years In the event of a justified claim Nickel Electro will replace any defective component free of charge This warranty does NOT apply if 1...

Page 8: ...of laboratory medical and clinical equipment Oldmixon Crescent Weston Super Mare North Somerset BS24 9BL United Kingdom Tel 44 0 1934 626691 Fax 44 0 1934 630300 Email clifton nickel electro co uk www...

Reviews: