7. CLEANING AND DISINFECTION
NOTE
Always disconnect the power cord before cleaning the system and while
unit is drying if wet-wiped.
Disinfection agents should be commercially available cleansers commonly used for
disinfection of electronic equipment in hospitals, such as ethyl or isopropyl alcohols, or
disinfecting sprays containing quaternary ammonium compounds or hydrogen peroxide.
Do not use strongly caustic or acidic cleansers such as “Clorox” hypochlorite bleach,
ammonia, muriatic acid or similar products. Do not use acetone, methyl ethyl ketone or
halogenated / chlorinated hydrocarbon solvents or cleaners containing any of these restricted
compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids over the
cabinet.
Do not allow liquids to enter the cabinet seams or ventilation openings.
8. MAINTENANCE, SERVICING AND REPAIR, WARRANTY
Performance of routine maintenance is essential to proper operation of the illuminator.
Follow the steps below for replacing filters and lamps.
Regular diagnostics can contribute to identifying potential problems before they become
serious; thus, enhancing the instrument’s reliability and extending its useful operating life.
Diagnostic services can be obtained from your local representative.
Defective items or equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer’s parts
only.
8.1
Filter Replacement
Every 250 operating hours the air filter must be replaced. The time for replacement is
notified by the system software (see paragraph 6.2.2.4.2). To replace the filter:
Slide open the filter drawer located under the lightsource (it is labeled
FILTER – PULL)
Remove and dispose of old filter
Obtain a new filter from the accessories kit and place into filter drawer
Close filter drawer
Reset filter message per paragraph 6.2.2.4.3