Skip weld collar to frame
Note
LoadRunner™ Series Installation & Service Manual
Leave access to easily install the bolt and nut that secures the load cell.
Figure 3-5. Skip Weld Collar to Frame
14. Apply grease to each center pivot using existing lubrication points.
15. Reinstall the load pin, lining up the bolt hole with the hole in the collar.
1/2-20 x 5
”
Screw
Load Pin
Important
Do NOT strike the load cell directly, as it may cause damage to the load cell. Use a wood block (or similar
material) to cushion the load cell, and tap lightly. If the load pin does not go in by hand, or with moderate
tapping, the holes need to be aligned better, or opened up with a grinder.
16. Insert bolt through the top of the collar and secure with nut at the bottom.
Figure 3-6. Insert Load Pin