4.5.6 Date/Time
The system time may be set under this menu. Touch “ZERO”
to adjust the value. Press “MODE” to shift to the next position.
Touch “LOG” to cancel. Press “MENU” to confirm and exit.
(Figure 4-5h)
Figure 4-5h
4.5.7 Password
The system password can be changed. First, enter the old
password, then enter the new password and confirm the
new password. The default System password is “123”. Touch
“ZERO” to adjust the value. Press “MODE” to shift to the next
position. Touch “LOG” to cancel; Push “MENU” to confirm and
exit. (Figure 4-5i)
Figure 4-5i
4.5.8 Key Setting
Set the default function of the “LOG” key from the home
screen. The function can be set to print or store the current
displayed value. Press “ZERO” or “MODE” to select the proper
setting. Press “LOG” to cancel; touch “MENU” to confirm and
exit. (Figure 4-5j)
Figure 4-5j
4.5.9 RS232 Baurate
Adjust Baurate to available bits per second selection in Figure
4-5k.
Figure 4-5k
4.5.10 Default Setting
If you make a selection that you feel has caused the gauge
to operate improperly, you can restore it to the factory default
settings. Carefully use this function! (Figure 4-5l)
Figure 4-5l
4.6 Language
Select between English, German and Chinese (Figure 4-6a)
Figure 4-6a
4.7 Calibration
Users can field-calibrate the gauge. First, enter the system pass-
word (Default is 123) by pressing the “ZERO” and “MODE” keys.
Then press “MENU” to confirm. (Figure 4-7a)
Figure 4-7a
1
Calibration Point
2
Current Value
3
Standard Input Value
Figure 4-7b
Load a standard force on the gauge. Wait a moment for the force
to stabilize. The current value (2) should equal the standard force
applied.
If the values do not match, press “ZERO” and “MODE” keys to
correct the standard input value (3).
Press “MENU” to enter the next calibration point. After any of the
calibration points have been completed, touch “LOG” to exit the
calibration mode. Then save the calibration or discard by pressing
“Yes” or “No”.
After completing the calibration of the 5th point, the confirmation
window will automatically ask to “Save and Exit” by selecting
“Yes” or “No”. (Figure 4-7c)
Figure 4-7c
Press “ZERO” or “MODE” to select, then press “MENU”. If “Yes”
is selected, “Calibration Complete!” is displayed.
6
www.
GlobalTestSupply
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com