DEVICE SETTINGS AND MAINTENANCE: AR/KM Measurement Accuracy Check
●
Values marked on the labels of the spherical model eye
VD value
Diopter (Unit: D)
Corneal curvature radius (Unit: mm)
• When the vertex distance is set to a value other than “12.00mm”, set the “2. VERTEX D.” parameter
to “12.00mm” before conducting AR measurement.
• Keep fingers off the lens surface of the spherical model eye. For severe stains, wipe the area with
gauze dampened with alcohol.