15
INTRODUCTION: Configuration and Functions
7
Auto shot icon
Indicates the setting of the auto shot function.
8
CYL mode
Indicates the selected cylinder mode (CYL+, CYL-, CYL±).
9
Indication icon
An icon indicating the device or measurement status is displayed at the top of the screen.
10
Minimum pupil diameter mark
If the pupil diameter is smaller than this mark or eyelashes are on this mark, measurement may not be
possible.
11
Mire ring
Used as an alignment reference ring.
12
Target
Used as a guide to locate the patient’s eye in the center of the screen. Align the target with the mire
ring projected on the patient’s eye.
When the mire ring is detected, the indication changes. (
)
13
Focusing indicator
Indicates the distance between the measuring unit and the patient’s eye.
Measurement starts automatically as soon as alignment and focusing become optimum.
(No icon)
Pressing the start button starts measurement.
Cataract measure-
ment mode icon
Indicates that the eye has been measured in cataract measurement mode.
Eye Care card icon
Displayed when an Eye Care card is inserted.
ID icon
Displayed when patient ID has been input with the barcode scanner or mag-
netic card reader.
When an Eye Care card is not inserted, patient ID is displayed instead.
Sagittal icon
Indicates that KM measurement has been performed and sagittal
measurement is possible.
Summary of Contents for ARK-1S
Page 1: ...AUTO REF KERATOMETER AUTO REF KERATOMETER ARK 1s ARK 1s OPERATOR S MANUAL OPERATOR S MANUAL ...
Page 6: ...IV ...
Page 76: ...70 OPERATING PROCEDURE Operation when Peripheral Devices are Connected ...
Page 112: ...106 SPECIFICATIONS AND TECHNICAL INFORMATION EMC Electromagnetic Compatibility ...