27
CHECK AND ADJUSTMENT PROCEDURES: Size Check and Adjustment
3.5
Size Check and Adjustment
Adjust size on the parameter so that the outer diameter of the lens flat edged using Circle 45 is within
the tolerance of
φ
45.0±0.1 mm.
3.5.1
Size check
1
Enter service mode from the LE-700 home screen.
“3.2.1 Entering service mode” (page 20)
2
Press the “Maintenance” tab.
The maintenance screen appears.
3
Press Circle 45
.
The processing condition screen appears.
4
Specify the flat edging conditions.
1)
Set each item as follows:
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Flat
“Soft”
→
None
“Polish”
→
None
“Layout”
→
Passive
“FC”
→
None
2)
Press
.
The layout screen appears.