32
CHECK AND ADJUSTMENT PROCEDURES: Size Check and Adjustment
5
Change the parameter value by the outer
diameter difference from 45.00 mm.
When the outer diameter is 45.20 mm in diameter,
reduce the parameter value by 0.20.
Enter a value with the numeric keypad and press
.
6
Press the “Maintenance” tab.
The maintenance screen appears.
7
Press Circle 45
.
The processing condition screen appears.
8
Specify the flat edging conditions.
1)
Set each item as follows:
“Material”
→
CR39
“Type”
→
Single
“Frame”
→
Flat
“Soft”
→
None
“Polish”
→
None
“Layout”
→
Passive
“FC”
→
None
2)
Press
.
The layout screen appears.
9
Continue the processing method on the layout screen using the same procedures
explained by the following steps starting from size setting.
to
10
Measure the outer diameter of the processed lens with a caliper to check it is in the
tolerance of
φ
45.0±0.1 mm.
11
to
until the outer diameter is in the tolerance of
φ
45.0±0.1
mm.