2
FOR SAFE INSTALLATION: During Installation
1.3
During Installation
WARNING
• Use a hand cart when moving the instrument. At least
two persons are necessary to lift the instrument.
Failure due to dropping of the instrument, back
injury, or stumbling may occur.
• Be sure to lift the instrument by holding it at the points
, ,
, and
at the bottom of both sides
(see the figure to the right). Do not lift the instrument
by its cover.
Failure to do so may cause the instrument to fall
resulting in injury or malfunction.
• Be careful not to get fingers caught when setting the
instrument down.
Fingers may get caught between the table and instrument resulting in injury.
• Set the instrument on a table whose area is larger (50 mm or more) than the outer dimensions of the
instrument as shown below. A table height of 650 to 700 mm is recommended to allow space for the
pump tank to be installed under the table. Additionally, the table must be sturdy enough to bear the
weight of the instrument (33 kg) and be provided with adjusters on its legs to allow it to be level and
stable.
Failure to do so could cause the instrument to fall in the event of impact or accidents.
An opening must be made in the table to pass the pipe through.
• Install the instrument in a stable and level place where vibration or shock does not occur.
Tilting or vibration can cause the instrument to fall or prevent normal lens processing.
B
A
D
C
26.5
316.5
490
543
50
3-
115
304
2-R50
R70
237
242
261
269
Position of instrument rubber foot (three spots)
Opening for drain pipes
Opening for cooling water and
cleaning water pipes
(unit: mm)
Instrument perimeter
Table perimeter (example)
A height of 650 to 700 mm
recommended