8 - 12
8.4 Replacing Fuses
If the instrument can not be activated though the power switch is turned ON, the fuses may be
blown. Replace the fuses with new ones.
CAUTION
• Be sure to turn OFF the power switch and pull out the power cable before replacing
fuses.
Otherwise, you may get an electrical shock.
• Use specified fuses only. (115 V AC
→
T10A 250V)
(230 V AC
→
T6.3A 250V)
Fuses other than the specified ones may cause a fire.
1. Turn OFF (
{
{
{
{
{
) the power switch and disconnect the power cable from the wall outlet.
2. Disconnect the power cable from the inlet.
3. Remove the fuse holder under the inlet.
Pull out the fuse holder by pressing the lever.
4 Remove the old fuses from the holder and
replace them with new ones.
5. Push the fuse holder into the body as before.
8.5 Cleaning the Exterior
When the exterior of the instrument such as the covers or panels become soiled, wipe them with
a soft cloth. For stubborn soiling, immerse a cloth in a neutral detergent, wring it well, and wipe
the area. Wipe off with a dry cloth to finish.
CAUTION
• Never use organic solvents such as paint thinner.
This could dissolve the surface.
Lever