8 - 4
Drain pipe
Feedwater hoses
Filter joint
Drain pipe connector
Drain cover
Stocking filter
8.2 Replacing the Water and Filter (115 V regions)
The procedure differs depending on whether the optional FP-100 is equipped or not.
8.2.1 Replacing the water (when the optional FP-100 is equipped)
Replace the processing water and stocking filter in the tank periodically. It is recommended that
the water be replaced for every 100 lenses processed.
NOTE
• Do not mix antifoam agents with the processing water since the filter at the bottom of
the tank will be clogged, which prevents filtering.
• When the “Please Clean Tank & Pump and Replace Filter” message appears and the
processing is stopped, replace the processing water and stocking filter, and remove the
waste accumulated in the drain pipe connector.
1. Take the tank out of the cabinet.
Open the cabinet and draw the tank toward you.
2. Disconnect the two feedwater hoses.
Press the red button on the hose to disconnect.
3. Disconnect the drain pipe from the tank.
4. Drop the stocking filter into the tank.
1) Remove the drain cover from the drain pipe
connector.
Drop the waste accumulated on top of the
drain cover into the tank.
2) Drop the stocking filter into the tank.
5. Remove the lid of the tank and put the waste in the stocking filter into the tank.
1) Remove the lid of the cover.
2) Rip the stocking filter with a cutter and remove the waste and put it into the tank.
Proper waste filtration cannot be performed with the waste in the stocking filter. Put the
ripped stocking filter into the tank, as well.