9 - 4
Whole system
Dimensions:
520 (W) × 484 (D) × 330 (H) mm (only main body)
Weight:
42 kg (only main body)
Power supply:
A. AC115 V 50/60 Hz
B. AC230 V 50/60 Hz
(Main supply voltage fluctuation should not exceed ±10% of the
nominal voltage.)
* Each setting of A and B is set in factory.
Power consumption:
1.2 k VA (max. during processing)
Acoustic noise:
Maximum 70 dB (A)
according to Clause 1.7.4 of Machinery Directive
Environmental condition (during transport/storage)
Temperature:
–25 - 70ºC
Humidity:
10 - 100%
Pressure :
700 - 1060 hPa
(The conditions above indicate the state of the instrument packed.)
Environmental condition (during use)
Setting location:
Indoor
Temperature:
5 - 40ºC
Humidity:
Relative humidity does not exceed 50% at a maximum temperature
of 40ºC.
Pressure:
700 - 1060 hPa
Altitude:
Up to 1000 m at sea level
Transient overvoltages according to installation categories
(OVERVOLTAGE CATEGORIES)
II
Pollution degree
2 (IEC60664)